Alfa Laval separator: alarm A85 (MT60 fault), water content shows only dashes
EL-0075 · Solved · rating 0 · works for 0 engineers · occurred 20 Sept 2024
Source: /cases/el-0075
Symptoms#
- Separator with a water content transducer. The transducer had problems for about two weeks.
- The transducer board was replaced with a new one. After that, alarm A85 (MT60 fault) came up.
- The old board was fitted back — same alarm.
- The capacitance reading was about 300 pF. The water content (%) showed only dashes, as if the transducer saw nothing.
Checked#
| Check | Result |
|---|---|
| New transducer board | Alarm A85 |
| Old transducer board refitted | Same alarm — the board was not the cause |
Diagnosis#
The sensor in the pipe was heavily fouled. It had been reported as checked, but it was not really cleaned.
Fix#
Took the sensor out of the pipe and cleaned it to bright metal.
Result#
The alarm cleared and the transducer worked again.